Meditech Debrillator Monitor can work both in Automated External Debrillator (AED) mode and Manual mode and record parameters of ECG, HR and SpO2. Features: ·
Works both in automated external debrillator (AED) and Manual modes · Records ECG, HR and SpO2 · It has voice and visual prompts for charge prompt and shockable rhythm · It has a high-resolution display and a thermal printer
Display:
|
7 screen(17.7cm)diagonally
|
Sweep speed:
|
25mm/ sec
|
Number of leads for ECG:
|
3-lead ECG cable or 5-lead ECG cable, paddles
|
Output:
|
Biphasic
|
Energy sequence:
|
Monitor display indicates both selected and delivered energy
|
Charge time:
|
<7sec with a new fully charged battery
|
Charge prompts:
|
Voice and visual prompts
|
Electrode impedance measurement range:
|
0-250Ω
|
Lead selection:
|
Displayed on monitor, paddles, I, II, III, AVR, AVL, AVF, V
|
ECG size:
|
0.25, 0.5, 1, 1.5, 2, 4 cm/mV display on monitor
|
Heart rate:
|
20-300bpm
|
Heart rate alarm:
|
On/Off displayed on monitor, user-selectable
|
Smart alarm:
|
Beeper/voice prompts indicate shockable rhythm
|
SpO2 measurement range:
|
30~100%, ±2% between 80%~90%, others ±5%
|
Alarm range for SpO2 measurement range:
|
User set high limit and low limit
|
Alarm accuracy for SpO2 measurement range:
|
±10% within setting values
|
Alarm time accuracy for SpO2 measurement range:
|
Less than 12secs
|
Functions of AED mode:
|
Auto analyse and charge x3 with programmable auto energy level selection, screen prompts, and voice prompts
|
Shockable rhythms of AED mode:
|
Ventricular fibrillation with amplitude >=200uV ventricular tachycardia with rates >=140bpm QRS complex wave duration >=140ms
|
Charge control of AED mode:
|
Control on device front panel, press key on paddle
|
Prompts of AED mode:
|
Voice and visual prompts
|
Energy selection of Manual mode:
|
Selectable at 2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, 360 joules
|
Synchronization of Manual mode:
|
Synchronizes defibrillator pulse to patient's R-wave. ”SYNC” message displayed on monitor
|
Printer paper:
|
50mm thermal paper
|
Paper speed:
|
12.5mm/sec, 25mm/sec, 50mm/sec. User-selectable 6-second delay
|
Printing method:
|
High-resolution, thermal print head
|
Print-out modes:
|
Manual or automatic, user-configurable
|
On/Off control:
|
Front panel and paddle
|
Automatic function:
|
9-seconds recording initiated by alarm activation or defibrillator charge or defibrillator discharge
|
Battery:
|
Rechargeable, Ni-MH battery, 12V
|
Battery life:
|
For a new, fully charged battery- 60 defibrillator discharges at maximum energy, or 3 hours minimum of continuous ECG monitoring. Additional parameters will effect operating time with different functions
|