Index2 Fluke Series Pulse Oximeter Tester

The Index 2 is a lightweight, portable tool that includes preloaded manufacturers’ R-curves and the ability to define other “makes" for testing pulse oximeters.Index2 Fluke Series Pulse Oximeter Tester

Motion presets, player mode, transmission level control (TLC), and computer commands boost testing ability. The Index 2 can also be configured to include an optional electrical simulation feature with probe test. Optical and electrical simulations allow technicians to isolate problems quickly. The probe test identifies defective probes with quantitative test results. 

 

Features

Portable

Tenpreloaded manufacturers’ R-curves

User-definable “makes for most other manufacturers

R-curves for Masimo, Nonin and Philips Medical Systems (formerly Agilent / H-P) oximeters

Six downloadable R-curve spaces available

Simultaneous simulation of motion and arterial-oxygen levels

Arterial wave-amplitude scale, calibrated in units of perfusion

Tap/shiver motion simulations to explore the impact of motion

RS-232 port for computer control

Physiological finger

 

Electrical simulations with probe testing

 

Specifications
O2
Range: 35 % to 100 %
Resolution: 1 %
Accuracy: 100 % to 75 %: ± 1 % ± accuracy of the device under test; 74 % to 50 %: ± 3 % ± the accuracy of the device under test; < 50 % unspecified
Repeatability: Repeatability: ± 1 standard deviation
Rate
Range: 30 BPM - 250 BPM
Resolution: 1 BPM
Accuracy: 1 % ± 1 BPM
Pulse Amplitude
Range: 0 to 100 % of nominal pulse amplitude
Resolution: 1 %
Probe test
Continuity/Resistance: Pulse amplitude is 20 % of maximum pass-through brightness
Test Matrix: Measures all combinations of possible interconnections in an XX point matrix
Range: 250 O to 150 kO
Accuracy: ± 5 % of reading
LED/Detector voltage test
Test Format: Measures the voltage drop across Red LED, infrared LED, and the photo detector when the internally generated test signal is applied
Test Signal: Constant current source @ 1 mA
Open Circuit: 2.5 V max
Measurement/Display Range: 0 V to 4 V
Accuracy: ± 5 % of reading, 0.4 V to 4 V
Dynamic Test
Test Format: Photodetector/diode response to both the red and infrared light generated by the probe when pulsed by an internal test signal
Test Signal: Pulsed between the two LEDs; constant current level @ 1 mA
Test Results: Nominal range of 0 to 2000
Checksum
  Sum of all locations in the program chip; for service use only
General Information
Display: 2-line x 24-character super twist LCD
Battery Life: At least 4 hours of continuous use
Battery Charger: Lead acid battery charger only
Input:100 V to 250 V, 50-60 Hz, 0.3 A
Output:+12 V dc, 0.5 A
Operating temperature: 15 oC to 35 oC (59 oF to 95 oF)
Storage temperature: 0 oC to 50 oC (32 oF to 122 oF)
Humidity 10% to 90% non-condensing
Dimensions (WxDxH): 26.4 cm x 25.4 cm x 10.2 cm
(10.4 in x 10 in x 4 in)
Weight: 2.38 kg (5 lb)